ref | title | DOI | material type | comment |
---|---|---|---|---|
810 | One-pot, direct glucose detection in human whole blood without using a dilution factor by a magnetic nanozyme with dual enzymatic activity | https://doi.org/10.1016/j.jallcom.2020.156012 | Metal | Glucose oxidase immobilized-Au nanoparticle attached-magnetic SiO2 (GOx@Au@MagSiO2) microspheres |
ref | material | size | size err | size unit | size type | size comment | BET | b nanozyme | b 10n | b unit | specific act | sa 10n | sa unit | comment |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
810 | GOx@Au@MagSiO2 | 6.5 | μm | SEM | The mean particle size and coefficient of variation for size distribution were calculated as 6.5 μm and 4.1%, respectively (Fig. 3A and Table 1). | 12.3 |
ref | material | enzyme type | substrate | pH | T | km | km err | km 10n | km unit | vmax | vmax err | vmax 10n | vmax unit | kcat | kcat err | kcat 10n | kcat unit | kcat/km | kcat/km
10n | kcat/km unit | comment |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
No matching records found |
ref | material | application | target | method | linear range | linear ran unit | LOD | lod unit | recovery | comment |
---|---|---|---|---|---|---|---|---|---|---|
810 | GOx@Au@MagSiO2 | One-pot, direct glucose detection | glucose | Color | 0.44–17.78 | mM | 0.66 | mM | The calibration plots obtained with the blood volume fractions of 0.20 and 0.46 mL/mL can be eventually used for direct determination of blood glucose concentration in the range of 0.5–7.0 mg/mL blood (2.8–38.9 mM). |